Prospective Randomized Crossover Study of a New Closed-loop Control System versus Pressure Support during Weaning from Mechanical Ventilation
Noémie Clavieras, Marc Wysocki, Yannaël Coisel, Fabrice Galia, Matthieu Conseil, Gerald Chanques, Boris Jung, Jean-Michel Arnal, Stefan Matecki, Nicolas Molinari, Samir Jaber

Prospective Randomized Crossover Study of a New Closed-loop Control System versus Pressure Support during Weaning from Mechanical Ventilation

Science & Mathematics Published 2013 Verification Pending

Overview & Synopsis

Background:: Intellivent is a new full closed-loop controlled ventilation that automatically adjusts both ventilation and oxygenation parameters. The authors compared gas exchange and breathing pattern variability of Intellivent and pressure support ventilation (PSV). Methods:: In a prospective, randomized, single-blind design crossover study, 14 patients were ventilated during the weaning phase, with Intellivent or PSV, for two periods of 24 h in a randomized order. Arterial blood gases were obtained after 1, 8, 16, and 24 h with each mode. Ventilatory parameters were recorded continuously in a breath-by-breath basis during the two study periods. The primary endpoint was oxygenation, estimated by the calculation of the difference between the Pao2/Fio2 ratio obtained after 24 h of ventilation and the Pao2/Fio2 ratio obtained at baseline in each mode. The variability in the ventilatory parameters was also evaluated by the coefficient of variation (SD to mean ratio). Results:: There were no adverse events or safety issues requiring premature interruption of both modes. The Pao2/Fio2 (mean ± SD) ratio improved significantly from 245 ± 75 at baseline to 294 ± 123 (P = 0.03) after 24 h of Intellivent. The coefficient of variation of inspiratory pressure and positive end-expiratory pressure (median [interquartile range]) were significantly higher with Intellivent, 16 [11–21] and 15 [7–23]%, compared with 6 [5–7] and 7 [5–10]% in PSV. Inspiratory pressure, positive end-expiratory pressure, and Fio2 changes were adjusted significantly more often with Intellivent compared with PSV. Conclusions:: Compared with PSV, Intellivent during a 24-h period improved the Pao2/Fio2 ratio in parallel with more variability in the ventilatory support and more changes in ventilation settings.

Indexed from repository: HAL Open Science (hal.science)
Loading document from verified source...
Keyboard Shortcuts: ← / → for pages • + / - for zoom • 0 to fit width Powered by Mozilla PDF.js v3.11

Document Information

Document Title Prospective Randomized Crossover Study of a New Closed-loop Control System versus Pressure Support during Weaning from Mechanical Ventilation
Author / Organization Noémie Clavieras, Marc Wysocki, Yannaël Coisel, Fabrice Galia, Matthieu Conseil, Gerald Chanques, Boris Jung, Jean-Michel Arnal, Stefan Matecki, Nicolas Molinari, Samir Jaber
Subject & Discipline Multidisciplinary Academic Open Research
Categories Science & Mathematics
Publication Year 2013
License / Rights Open Access / Creative Commons
Resource Verification Listed PDF (Verification Pending)
Original Host Website HAL Open Science (hal.science)
File Format application/pdf (Portable Document Format)
Language EN
Catalog ID DOC-190239
Original Source Link Visit Source Page on hal.science »